Z-0574-2015 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-0574-2015 was initiated by Zimmer, Inc. on Aug 26, 2013 and is designated Class II. Reason for recall: Possibility of the weld to the quick-connect breaking when force is applied to the handle for pedicle screw removal. A broken weld could lead to the use of an alternate method of screw removal… The recall status is terminated (terminated Feb 24, 2015). Affected quantity: 14 devices.

Recall Details

Product Type
Devices
Report Date
Dec 17, 2014
Initiation Date
Aug 26, 2013
Termination Date
Feb 24, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14 devices

Product Description

Zimmer Title 2 Bone Screw Remover Instrument, Part Number ET1093-01 Rx Only, Non-Sterile. The Screw Remover is a component of the Endius TITLE 2 Polyaxial Spinal System, used for posterior lumbar fusion procedures.

Reason for Recall

Possibility of the weld to the quick-connect breaking when force is applied to the handle for pedicle screw removal. A broken weld could lead to the use of an alternate method of screw removal, which could delay surgery up to 30 minutes.

Distribution Pattern

US nationwide distribution.

Code Information

Lot: L75066