Z-0580-2022 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- February 16, 2022
- Initiation Date
- January 7, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2,240 units
Product Description
Baxter CAPD System Transfer Set (CAPD MiniCap Transfer Set II (34cm)), Code No. T5C4484
Reason for Recall
Cleaning products, containing, but not limited to, hydrogen peroxide, bleach, alcohol, or antiseptic agents and solvent, such as those containing acetone, toluene, xylene, or cyclohexanone may cause damage, such as leaking or cracking, to the MiniCap Extended Life PD transfer sets.
Distribution Pattern
Worldwide
Code Information
All lot numbers within expiry