Z-0586-2014 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-0586-2014 was initiated by Zimmer, Inc. on Nov 5, 2013 and is designated Class II. Reason for recall: Based on a review of historic packaging validations, it was determined that a subset of product packaged at a Zimmer facility required testing to ensure that previously packaged product from this… The recall status is terminated (terminated May 5, 2015). Affected quantity: 42, 065 units total.

Recall Details

Product Type
Devices
Report Date
Jan 8, 2014
Initiation Date
Nov 5, 2013
Termination Date
May 5, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
42, 065 units total

Product Description

TIBIO FEM/FEM NAIL.

Reason for Recall

Based on a review of historic packaging validations, it was determined that a subset of product packaged at a Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. This testing determined that a packaging tray configuration which consists of an inner and outer sterile tray, resulted in failure

Distribution Pattern

Worldwide Distribution.

Code Information

Item Numbers 32-8553-311-11 32-8553-311-13 32-8553-311-15 32-8553-311-17 All lots with an expiration date prior to July 24, 2023