FDA device recall Z-0586-2017 was initiated by Roche Diabetes Care, Inc. on Sep 14, 2016 and is designated Class II. Reason for recall: Roche Diabetes Care, Inc. initiated a voluntary recall for the version 1.2.0 of Accu-Chek Connect Diabetes Management App (iOS), released on July 11, 2016, due to the software containing a programming… The recall status is terminated (terminated Jun 6, 2017). Affected quantity: 7,909 downloads of iOS version 1.2.0 of the Accu-Chek Connect Diabetes Management App in the US and 8,775 downloads…

Recall Details

Product Type
Devices
Report Date
Nov 23, 2016
Initiation Date
Sep 14, 2016
Termination Date
Jun 6, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
7,909 downloads of iOS version 1.2.0 of the Accu-Chek Connect Diabetes Management App in the US and 8,775 downloads Internationally

Product Description

Version 1.2.0 of Accu-Chek Connect Diabetes Management App (iOS) released on July 11, 2016

Reason for Recall

Roche Diabetes Care, Inc. initiated a voluntary recall for the version 1.2.0 of Accu-Chek Connect Diabetes Management App (iOS), released on July 11, 2016, due to the software containing a programming error (bug) in the Bolus Advisor feature, which could lead to an incorrect insulin bolus recommendation.

Distribution Pattern

US Distribution to the states of : Alabama, Arizona, Arkansas, California, Colorado, Connecticut, DC, Delaware, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Virginia, Washington, West Virginia, Wisconsin Foreign: None provided

Code Information

Version 1.2.0 of Accu-Chek Connect Diabetes Management App (iOS) released on July 11, 2016; Catalog number 07562462001 / GTIN number 00365702700000

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0586-2017
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.