Z-0589-2021 Class II Terminated

Recalled by RAYSEARCH LABORATORIES AB — Stockholm, N/A

FDA device recall Z-0589-2021 was initiated by RAYSEARCH LABORATORIES AB on November 11, 2020 and is designated Class II. Reason for recall: Two issues were found with the treatment plan report: 1) There may be incorrect information for the percent of a region of interest with a clinical goal that is outside the dose grid; and 2) For RaySt… The recall status is terminated (terminated May 9, 2023). Affected quantity: 124.

Recall Details

Product Type
Devices
Report Date
December 23, 2020
Initiation Date
November 11, 2020
Termination Date
May 9, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
124

Product Description

RayStation 4, 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, RayPlan 1, 2, 7, 8A, 8B, 9A, 9B, including all service packs. RayStation 10A and RayPlan 10A are also affected, but not the 10A service pack. (Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0, 9.1, 9.2, 10.0)

Reason for Recall

Two issues were found with the treatment plan report: 1) There may be incorrect information for the percent of a region of interest with a clinical goal that is outside the dose grid; and 2) For RayStation and RayPlan 9B and 10A, 6D couch angles may be incorrect.

Distribution Pattern

US Nationwide distribution.

Code Information

Software version numbers: 4.0.0.14 , 4.0.3.4 , 4.5.1.14 , 4.7.2.5 , 4.7.3.13 , 4.7.4.4 , 4.7.5.4 , 4.7.6.7 , 4.9.0.42 ,; 5.0.1.11 , 5.0.2.35 , 5.0.3.17 , 6.0.0.24 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.61 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154