Z-1103-2026 Class II Ongoing

Recalled by RAYSEARCH LABORATORIES AB — Stockholm

Recall Details

Product Type
Devices
Report Date
January 28, 2026
Initiation Date
November 28, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
10 units

Product Description

RayStation/RayPlan. Model/Catalog Numbers: 11.0.0, 11.0.1, 11.0.3, 11.0.4. Software Version: RayStation 11A, RayStation 11A SP1, RayStation 11A SP2, RayStation 11A SP3. Radiation Therapy Treatment Planning System

Reason for Recall

Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.

Distribution Pattern

Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS

Code Information

Lot Code: UDI: 0735000201038920210518, 0735000201043320210610, 0735000201044020210916, 35000201063120220616. GTIN: 07350002010389, 07350002010433, 07350002010440, 07350002010631. Serial Numbers: 11.0.0.951, 11.0.1.29, 11.0.3.116, 11.0.4.15. Software Revision: RayStation 11A, 11A SP1, 11A SP2, 11A SP3. Expiration Date: 2027-06-22.