Z-0595-2019 Class II Terminated

Recalled by Angiodynamics Inc. (Navilyst Medical Inc.) — Glens Falls, NY

FDA device recall Z-0595-2019 was initiated by Angiodynamics Inc. (Navilyst Medical Inc.) on Oct 16, 2018 and is designated Class II. Reason for recall: Specific lots of Xcela Power Injectable and BioFlo PICCs are labelled with incorrect expiration dates. The labelled expiration date extends the shelf life of the product beyond the date supported by… The recall status is terminated (terminated Jun 26, 2020). Affected quantity: Box of 1 Units.

Recall Details

Product Type
Devices
Report Date
Dec 19, 2018
Initiation Date
Oct 16, 2018
Termination Date
Jun 26, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Box of 1 Units

Product Description

Xcela Power Injectable PICC with PASV (XCELA PASV 4F SL 55CM Catheter Kit CE PG), Cat no. 55-711

Reason for Recall

Specific lots of Xcela Power Injectable and BioFlo PICCs are labelled with incorrect expiration dates. The labelled expiration date extends the shelf life of the product beyond the date supported by validation testing. Although the product under recall has not yet reached its validated expiry date, removal of the affected product is warranted to prevent use beyond the validated expiry date.

Distribution Pattern

Distributed to accounts in Florida. Foreign distribution to Great Britain.

Code Information

Lot 5096448

Source

Dataset
openFDA device recall enforcement reports