Z-0601-2019 Class III Terminated
Recall Details
- Product Type
- Devices
- Report Date
- December 19, 2018
- Initiation Date
- October 5, 2018
- Termination Date
- June 16, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1000
Product Description
Hemoglobin A1c Control, Part no. 220232 for use as quality control material to monitor the precision of laboratory testing procedures for HbA1c quantitation.
Reason for Recall
Tosoh Bioscience has become aware of potential degradation of HbA1c control lot # 7055 before the 60-days product stability claim. If degradation of these controls occurs, P00 peaks may be seen and the % HbA1c control values may drift higher than the assigned range. This may result in delays when performing routine instrument quality control. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received two (2) complaints related to this issue with no serious injuries reported.
Distribution Pattern
Distributed nationwide. Foreign distribution to Canada.
Code Information
Lot 7055