Z-2968-2020 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- September 23, 2020
- Initiation Date
- August 7, 2020
- Termination Date
- June 21, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 39 boxes
Product Description
Tosoh Bioscience AIA-PACK BHCG Calibration Verification Test Set-In Vitro Diagnostic for BHCG-pregnancy Part number: 020661 The AIA-Pack BHCG Calibrator Set is intended for IN VITRO DIAGNOSTIC USE ONLY for the calibration of the ST AIA-PACK BHCG assay
Reason for Recall
Decreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up to 20%.
Distribution Pattern
Worldwide Distribution: US (Nationwide) and countries of: Bogota, Cayman Islands, Mexico, Peru, and San Salvador.
Code Information
Lot Numbers: J803507, JX03508, JZ03509, A203510, A403511