Z-0625-2016 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

FDA device recall Z-0625-2016 was initiated by Zimmer Biomet, Inc. on Nov 17, 2015 and is designated Class II. Reason for recall: Zimmer Biomet is initiating a recall of the Virage Occipital Eyelet, as it was found through product testing that the staking of the set screw was out of specification and can interfere with the… The recall status is terminated (terminated Aug 12, 2016). Affected quantity: 372.

Recall Details

Product Type
Devices
Report Date
Jan 20, 2016
Initiation Date
Nov 17, 2015
Termination Date
Aug 12, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
372

Product Description

Zimmer, Virage OCT Spinal Fixation System, Item Number: 07.01738.001, a posterior system intended to achieve an implant and construct of the Occipital-Cervical-Thoracic spine (Occiput-T3). UDI Number 889024328952

Reason for Recall

Zimmer Biomet is initiating a recall of the Virage Occipital Eyelet, as it was found through product testing that the staking of the set screw was out of specification and can interfere with the insertion of the occipital rod when the set screw is staked too low.

Distribution Pattern

US: AL, AZ, CA, CO, FL, GA, ID, MI, MN, MO, NV, NJ, OK, OR, PA, SD, TN, TX, WA, WI. OUS: AUSTRALIA, GERMANY, SWITZERLAND, SWEDEN.

Code Information

Lots: 62968065, 62947660, 62536230, 62745662, 62773482, 62758613, 62874869

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0625-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.