Z-0625-2019 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- December 26, 2018
- Initiation Date
- November 15, 2018
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
Product Description
Covidien Emprint Long Percutaneous Antenna with Thermosphere Technology, Material CA30L2 intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.
Reason for Recall
The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
Distribution Pattern
US Nationwide
Code Information