Z-0630-2022 Class II Ongoing

Recalled by Becton Dickinson & Co. — Sparks, MD

Recall Details

Product Type
Devices
Report Date
February 16, 2022
Initiation Date
January 5, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
126,952 Analyzers Updated:129, 848 Analyzers

Product Description

BD Veritor Plus Analyzer - intended to provide rapid test results in near-patient settings. Catalog Number: 256066

Reason for Recall

May have the potential to overheat and/or cause fire, the issue can occur when an unauthorized alternative AC power adapter, and not the one provided by BD is used to power the instrument.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Armenia, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Dominican Republic, Ecuador, Fiji, Guatemala, Hong Kong, Honduras, Haiti, Indonesia, India, Jamaica, Cambodia, South Korea, Macao, Mexico, New Zealand, Panama, Peru, Philippines, Singapore, El Salvador, Thailand, Trinidad and Tobago, Taiwan, Uruguay, Vietnam.

Code Information

All unexpired analyzers (Exp. Date 12/2021 - 8/2024) UDI: 00382902560661