Z-0646-2016 Class II Terminated

Recalled by Zimmer Biomet, Inc. — Warsaw, IN

FDA device recall Z-0646-2016 was initiated by Zimmer Biomet, Inc. on Dec 9, 2015 and is designated Class II. Reason for recall: Product on the package label did not match what was inside the package. The label was for part number 00-2359-058-35 (Zimmer Periarticular Locking Plate System 3.5mm Locking Screw 58mm Length) lot… The recall status is terminated (terminated Nov 8, 2016). Affected quantity: 400.

Recall Details

Product Type
Devices
Report Date
Jan 20, 2016
Initiation Date
Dec 9, 2015
Termination Date
Nov 8, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
400

Product Description

Zimmer Periarticular Locking Plate System 3.5mm Locking Screw 58 mm Length; Zimmer Periarticular Locking Plate System 2.4mm Locking Screw 14 mm Length. Indicated for temporary internal fixation and stabilization of osteotomies and fractures.

Reason for Recall

Product on the package label did not match what was inside the package. The label was for part number 00-2359-058-35 (Zimmer Periarticular Locking Plate System 3.5mm Locking Screw 58mm Length) lot number 62918019 while the product inside the package was part number 00-2359-014-24 (Zimmer Periarticular Locking Plate System 2.4mm Locking Screw 14mm Length) lot number 62896038.

Distribution Pattern

Worldwide Distribution-US including the states of CA, NY, PA, AL, FL, GA, ID, WI, MO, SC, NC, and ME, and the countries of Brazil, Dominican Republic, Taiwan, China, Uruguay, Australia, Japan, Malaysia, India, Belgium, France, Italy, Netherlands, Poland, United Arab Emirates, and Canada.

Code Information

Item 00-2359-058-35, Lot 62918019 PERI LOCKING SCREW 3.5MM Item 00-2359-014-24, Lot 62896038 PERI LOCKING SCREW 2.4MM

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0646-2016
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.