Z-0660-2013 Class II Terminated
FDA device recall Z-0660-2013 was initiated by Biomet, Inc. on Dec 12, 2006 and is designated Class II. Reason for recall: Biomet is recalling Part Number PT-113950 PT Hybrid Glenoid Post, following an investigation which identified that the male thread of the post may be oversized. This oversized condition can vary in… The recall status is terminated (terminated Nov 8, 2013). Affected quantity: 1050.
Recall Details
- Product Type
- Devices
- Report Date
- Jan 16, 2013
- Initiation Date
- Dec 12, 2006
- Termination Date
- Nov 8, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1050
Product Description
PT113950 PT HYBRID Glenoid Post Regenerex porous titanium construct Sterile R Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
Reason for Recall
Biomet is recalling Part Number PT-113950 PT Hybrid Glenoid Post, following an investigation which identified that the male thread of the post may be oversized. This oversized condition can vary in degree and may lead to the following three events: 1) If the PT Hybrid Glenoid Post is not fully seated into the Hybrid Base and the implant construct is implanted, then a gap of 1-3 mm will be prese
Distribution Pattern
Worldwide Distribution -- US (nationwide) including the countries of , Canada, Japan, Argentina, Australia ,Netherlands, Europe and Chile.
Code Information
Part 113950 Lot 093540, 093550,093570,093580,093590,093600,093610,093620,093720, 093730,093740,093760,093770,093780,282590,282600,282620,282630, 282680,282700,282710,489160,489180,489200,489210,489770,489780, 489790,714480,714550,714570,714700,714730,722210,722250,722280, 984770,984780,984790,984800,984810,984820,984830,984840, and 984850