Z-0665-2022 Class III Ongoing

Recalled by Becton Dickinson & Company — Franklin Lakes, NJ

Recall Details

Product Type
Devices
Report Date
March 2, 2022
Initiation Date
January 6, 2022
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
288,152,790 Units

Product Description

BD Vacutainer Luer Lok Access Device - intended as a sterile, single use device designed to provide a connection to a catheter port equipped with a threaded female luer during blood or urine sample collection using evacuated BD Vacutainer Blood or Urine Collection Tubes Catalog Number: 364902, 36490200

Reason for Recall

May shed contaminant particles into the urine specimen that may be misidentified as Red Blood Cells when used in conjunction with the Sysmex UF-1000i or UX-2000 Analyzers and lead to false elevations in RBC counts

Distribution Pattern

US Nationwide distribution in the state of IL.

Code Information

All Lots within Expiry Catalog No: 364902 UDI: (01)50382903649025 Catalog Number: 36490200 UDI:(01)60382903649022