Z-0665-2022 Class III Ongoing

Recalled by Becton Dickinson & Company — Franklin Lakes, NJ

FDA device recall Z-0665-2022 was initiated by Becton Dickinson & Company on Jan 6, 2022 and is designated Class III. Reason for recall: May shed contaminant particles into the urine specimen that may be misidentified as Red Blood Cells when used in conjunction with the Sysmex UF-1000i or UX-2000 Analyzers and lead to false elevations… The recall status is ongoing. Affected quantity: 288,152,790 Units.

Recall Details

Product Type
Devices
Report Date
Mar 2, 2022
Initiation Date
Jan 6, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
288,152,790 Units

Product Description

BD Vacutainer Luer Lok Access Device - intended as a sterile, single use device designed to provide a connection to a catheter port equipped with a threaded female luer during blood or urine sample collection using evacuated BD Vacutainer Blood or Urine Collection Tubes Catalog Number: 364902, 36490200

Reason for Recall

May shed contaminant particles into the urine specimen that may be misidentified as Red Blood Cells when used in conjunction with the Sysmex UF-1000i or UX-2000 Analyzers and lead to false elevations in RBC counts

Distribution Pattern

US Nationwide distribution in the state of IL.

Code Information

All Lots within Expiry Catalog No: 364902 UDI: (01)50382903649025 Catalog Number: 36490200 UDI:(01)60382903649022

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0665-2022
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026