Z-0670-2022 Class II Terminated

Recalled by Biomet, Inc. — Warsaw, IN

FDA device recall Z-0670-2022 was initiated by Biomet, Inc. on Jan 7, 2022 and is designated Class II. Reason for recall: The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was… The recall status is terminated (terminated Apr 13, 2023). Affected quantity: 19 units.

Recall Details

Product Type
Devices
Report Date
Mar 2, 2022
Initiation Date
Jan 7, 2022
Termination Date
Apr 13, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
19 units

Product Description

Arcos Modular Revision Hip System, Standard Cone Prox Body, Porous Plasma,Size A, 60 MM Item Number: 11-301301

Reason for Recall

The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Ecuador, Netherlands.

Code Information

Lot Number: 853650, 853810 UDI: (01)00880304468030(17) 290122(10)853650; (01)00880304468030(17) 290211(10)853810

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0670-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 24, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.