Z-0675-2022 Class II Terminated

Recalled by Biomet, Inc. — Warsaw, IN

Recall Details

Product Type
Devices
Report Date
March 2, 2022
Initiation Date
January 7, 2022
Termination Date
April 13, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12 units

Product Description

Comprehensive Shoulder System Mini Humeral Stem, Porous Plasma, 12 MM, 83 MM Long Item Number: 113632

Reason for Recall

The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, Ecuador, Netherlands.

Code Information

Lot Number: 327160 UDI: (01)00880304449749(17)290819(10)327160