Z-0681-2018 Class II Terminated

Recalled by Vygon MFG, Inc., dba/ Churchill Medical Systems, Inc. — Dover, NH

Recall Details

Product Type
Devices
Report Date
February 28, 2018
Initiation Date
July 24, 2017
Termination Date
November 7, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
156

Product Description

19Ga x 1 Huber Needle Set. Product Code: MBS-1910.

Reason for Recall

Vygon is conducting a voluntary recall of the products and lots listed in this recall, being initiated because our Tyvek supplier has notified Vygon of issues with their product, specifically the Tyvek may not be coated completely causing gaps in the adhesive used to seal the product packaging. This potential lack of adhesive may affect the sterility of the product.

Distribution Pattern

U.S.

Code Information

Lot Number: 1706049D