Z-0678-2018 Class II Terminated
FDA device recall Z-0678-2018 was initiated by Vygon MFG, Inc., Dba/ Churchill Medical Systems, Inc. on Jul 24, 2017 and is designated Class II. Reason for recall: Vygon is conducting a voluntary recall of the products and lots listed in this recall, being initiated because our Tyvek supplier has notified Vygon of issues with their product, specifically the… The recall status is terminated (terminated Nov 7, 2019).
Recall details
- Product Type
- Devices
- Report Date
- Feb 28, 2018
- Initiation Date
- Jul 24, 2017
- Termination Date
- Nov 7, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
Product description
Nutrisafe 2 35 mL Syringe for Safety Connection. Product Code: 1015.352M
Reason for recall
Vygon is conducting a voluntary recall of the products and lots listed in this recall, being initiated because our Tyvek supplier has notified Vygon of issues with their product, specifically the Tyvek may not be coated completely causing gaps in the adhesive used to seal the product packaging. This potential lack of adhesive may affect the sterility of the product.
Distribution pattern
U.S.
Code information
Lot Number: 1705060D
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0678-2018