Z-0694-2019 Class II Terminated

Recalled by Philips Medical Systems Nederlands — Best

FDA device recall Z-0694-2019 was initiated by Philips Medical Systems Nederlands on Dec 3, 2018 and is designated Class II. Reason for recall: During automatic tabletop movement, pressing and holding the Stop Table button may lead to an unexpected restart of the table movement. The recall status is terminated (terminated May 12, 2020). Affected quantity: 35.

Recall Details

Product Type
Devices
Report Date
Jan 16, 2019
Initiation Date
Dec 3, 2018
Termination Date
May 12, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
35

Product Description

Ingenia Elition S, Ingenia Elition X

Reason for Recall

During automatic tabletop movement, pressing and holding the Stop Table button may lead to an unexpected restart of the table movement.

Distribution Pattern

CA, MN, NJ, NY, OH, TN, TX. Foreign distribution to Canada, Belgium, France, Germany, INdia, Japna, Netherlands, South Africa, Sweden, Switzerland, UAE.

Code Information

Software version R5.5.0 Serial numbers: 45027 45011 45010 45024 45032 45026 45054 45035 45051 45052 45007 45043 45044 45031 45039 45025 46001 45033 45050 45034 45049 45037 45048 46004 45046 45008 45058 45028 45029 45030 45042 46005 46011 45040 45045