Z-0700-2017 Class II Terminated

Recalled by Medtronic Neuromodulation — Minneapolis, MN

FDA device recall Z-0700-2017 was initiated by Medtronic Neuromodulation on Oct 3, 2016 and is designated Class II. Reason for recall: Medtronic has identified an increased number of complaints from customers involving reports of Rechargers (Medtronic Model 37751 Recharger, which is included in Models 37754 and 97754 Charging… The recall status is terminated (terminated May 4, 2022). Affected quantity: 42,887.

Recall Details

Product Type
Devices
Report Date
Dec 14, 2016
Initiation Date
Oct 3, 2016
Termination Date
May 4, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
42,887

Product Description

Model 37751 Recharger Product Usage: The Medtronic 37751 Recharger is a hand-held device used to charge the battery in a patient s neurostimulator. It includes a display to provide information on the charging system. The charging system consists of the Model 37751 Recharger, an AC power supply and power cord (Model 37761) and an antenna.

Reason for Recall

Medtronic has identified an increased number of complaints from customers involving reports of Rechargers (Medtronic Model 37751 Recharger, which is included in Models 37754 and 97754 Charging Systems,Spinal Cord Stimulation)that are in an unresponsive error state, where the Recharger is non-functional with a blank display screen and is beeping every 5 seconds. Medtronic has determined all Rechargers manufactured starting in November 2014 (indicated by serial numbers beginning with NKA4 or NKU4) are more susceptible to this error state.

Distribution Pattern

Worldwide Distribution - US Nationwide including DC OUS: Canada and others.

Code Information

serial numbers beginning with NKA4 or NKU4

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0700-2017
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.