Z-0701-2019 Class III Ongoing

Recalled by Boston Scientific Corporation — Marlborough, MA

Recall Details

Product Type
Devices
Report Date
January 16, 2019
Initiation Date
November 28, 2018
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,282 (all three types in Japan)

Product Description

IntellaNav MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or older, when used with a compatible mapping system

Reason for Recall

Supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.

Distribution Pattern

Japan

Code Information

IntellaNav Open-Irrigated Ablation Catheter Material Number: M004R96200 Batch: 22208397 Expiration Date: 6/4/2021 Batch: 22389516 Expiration Date: 7/16/2021 Batch: 22525798 Expiration Date: 2/14/2020