Z-0701-2019 Class III Ongoing

Recalled by Boston Scientific Corporation — Marlborough, MA

FDA device recall Z-0701-2019 was initiated by Boston Scientific Corporation on Nov 28, 2018 and is designated Class III. Reason for recall: Supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s… The recall status is ongoing. Affected quantity: 1,282 (all three types in Japan).

Recall Details

Product Type
Devices
Report Date
Jan 16, 2019
Initiation Date
Nov 28, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,282 (all three types in Japan)

Product Description

IntellaNav MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or older, when used with a compatible mapping system

Reason for Recall

Supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.

Distribution Pattern

Japan

Code Information

IntellaNav Open-Irrigated Ablation Catheter Material Number: M004R96200 Batch: 22208397 Expiration Date: 6/4/2021 Batch: 22389516 Expiration Date: 7/16/2021 Batch: 22525798 Expiration Date: 2/14/2020

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0701-2019
openFDA file date
Oct 21, 2026
Loaded by RegDataLab
Oct 23, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.