Z-0707-2021 Class II Ongoing

Recalled by ICU Medical, Inc. — San Clemente, CA

FDA device recall Z-0707-2021 was initiated by ICU Medical, Inc. on Aug 31, 2020 and is designated Class II. Reason for recall: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous… The recall status is ongoing. Affected quantity: 88700.

Recall details

Product Type
Devices
Report Date
Dec 30, 2020
Initiation Date
Aug 31, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
88700

Product description

31" (78 cm) Appx 3.4 ml, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros, Bag Hanger, REF: 011-CL3011; 40" (102 cm) Appx 4.9 ml, Admin Set w/ChemoLock, 20 Drop In-Line Drip Chamber, Spiros, Bag Hanger, REF: 011-CL3020; ChemoLock Port w/Spiros, Red Cap, REF: CL-35; 31" (79 cm) Appx 3.4 ml, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros, Purple Cap, Hanger, REF: CL3951; 40" (102 cm) Appx 7.7 ml, Admin Set w/ChemoLock w/Red Cap, 20 Drop In-Line Drip Chamber, 0.2 Micron Filter, Spiros w/Red Cap, Bag Hanger, REF: CL3962; 31" (79 cm) Appx 3.5 ml, Admin Set w/ChemoLock Bag Spike with Additive Port, 20 Drop In-Line Chamber w/15 Micron Filter, Spiros, Purple Cap, Hanger, REF: CL3999

Reason for recall

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Distribution pattern

Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.

Code information

Lots: 4867937, 4751234, 4799262, 4894789, 4787230, 4867947, 4897075, 4720480, 4735684, 4758456, 4775068, 4794504, 4860043, 4799897, 4787234, 4808045, 4860044 ,

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0707-2021
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for ICU Medical, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for ICU Medical, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about ICU Medical, Inc. →

Source: openFDA record for Z-0707-2021. Report a data issue: hello@regdatalab.com.