Z-2313-2026 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- June 17, 2026
- Initiation Date
- April 29, 2026
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 13,613
Product Description
Plum Duo Precision IV Pump, 40002-0403
Reason for Recall
Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of ID, IL, IN, KS, MA, MD, NM, NY, PA, TN, TX and the country of Canada.
Code Information
UDI-DI M335400031 Plum Duo Precision IV Pump v1.2.3