Z-0725-2017 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- December 14, 2016
- Initiation Date
- April 14, 2016
- Termination Date
- September 19, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15
Product Description
Magnesium on RX Imola analyser IVD
Reason for Recall
According to the firm, Carry over was observed when the amylase or pancreatic amylase test is run directly before or after the Magnesium assay. An update is required to the RX Imola analyser running order and an acid wash recommended when setting up the magnesium assay. A correction was made to the IFU for the Magnesium assay and all RX Imola customers in the USA were contacted with the updated instruction.
Distribution Pattern
Nationwide Distribution The product was only distributed to US Consignees.
Code Information
reagent: MG3880 analyser: RX4900