Z-2084-2019 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- July 31, 2019
- Initiation Date
- June 24, 2019
- Termination Date
- May 21, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 27
Product Description
RX Imola, Model Nos. RX4900
Reason for Recall
There was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument.
Distribution Pattern
Domestic distribution to CA CT MA MD MO NC NJ NY OH PA UT WV and PR. Worldwide foreign distribution.
Code Information
All serial numbers