Z-0728-2022 Class II Ongoing
FDA device recall Z-0728-2022 was initiated by Stryker Corporation on Jan 14, 2022 and is designated Class II. Reason for recall: A software defect in the camera control unit (CCU) will cause the image on the monitor to flip upside-down into an incorrect orientation. The potential of harms are conversion to open procedure… The recall status is ongoing. Affected quantity: 11,955.
Recall Details
- Product Type
- Devices
- Report Date
- Mar 16, 2022
- Initiation Date
- Jan 14, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 11,955
Product Description
1688 Camera Control Unit (CCU), Catalog number:1688010000. Used in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery wherever a laparoscope/endoscope/arthroscope/sinuscope is used.
Reason for Recall
A software defect in the camera control unit (CCU) will cause the image on the monitor to flip upside-down into an incorrect orientation. The potential of harms are conversion to open procedure, additional medical intervention, or a revision surgery.
Distribution Pattern
Worldwide distribution. US nationwide including Puerto Rico. OUS: BR, AE, NL, IN, FR, ES, HK, IT, GB, CO, PE, MY, TH, CA, JP, MX, AR, NP, SG, PH, ZA, CL, KR, CN
Code Information
All serial numbers of 1688 CCU, UDI: 07613327420081, with Software Revisions: 3.0.6, 4.0.13, and 4.0.18
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0728-2022
- openFDA file date
- Oct 26, 2026
- Loaded by RegDataLab