Z-0730-2022 Class II Ongoing

Recalled by Siemens Medical Solutions USA, Inc — Malvern, PA

FDA device recall Z-0730-2022 was initiated by Siemens Medical Solutions USA, Inc on Mar 7, 2022 and is designated Class II. Reason for recall: For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray… The recall status is ongoing. Affected quantity: 53 units.

Recall Details

Product Type
Devices
Report Date
Apr 13, 2022
Initiation Date
Mar 7, 2022
Voluntary/Mandated
FDA Mandated
Product Quantity
53 units

Product Description

YSIO X.Pree

Reason for Recall

For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray image, the x-ray collimator is automatically positioned in a way that the subsequent series of acquisition covers the defined field of view needed for each step. However, during the acquisition the collimation area displayed to the operator on the User Interface does not correctly represent the collimation area specified by the system. It indicates to the user an open collimator instead, e.g., the information displayed on the User Interface shows wider area of collimation than values preset prior to the examination. However, the collimation of the x-ray is performed correctly and always matches the examination area predefined by the user.

Distribution Pattern

US Nationwide, Worldwide distribution

Code Information

11107464

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0730-2022