Z-0739-2022 Class II Terminated

Recalled by Nextremity Solutions — Warsaw, IN

FDA device recall Z-0739-2022 was initiated by Nextremity Solutions on Jan 13, 2022 and is designated Class II. Reason for recall: During insertion of the middle phalanx implant into bone, the driver may deform the implant locking mechanism and result in inadequate mating force between the middle and proximal phalanx implants… The recall status is terminated (terminated May 24, 2024). Affected quantity: 23 units.

Recall details

Product Type
Devices
Report Date
Mar 16, 2022
Initiation Date
Jan 13, 2022
Termination Date
May 24, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
23 units

Product description

Nextra CH Cannulated Hammertoe Mini Instrument Kit-indicated for small bone reconstruction limited to inter-digital repair and fusion of the lesser toes SKU: CH-MINI-KT

Reason for recall

During insertion of the middle phalanx implant into bone, the driver may deform the implant locking mechanism and result in inadequate mating force between the middle and proximal phalanx implants, resulting in surgical delay, in-situ removal of the implant, use of alternative surgical approach, disassembly of the implants post-operatively, and revision

Distribution pattern

US Nationwide distribution in the states of AZ, IL, GA, LA, MD, NY, PA, TX, WA.

Code information

Lot Number: 1578401 UDI: 00817701026389

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0739-2022
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Nextremity Solutions? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0739-2022. Report a data issue: hello@regdatalab.com.