Z-0913-2022 Class II Terminated

Recalled by Nextremity Solutions — Warsaw, IN

FDA device recall Z-0913-2022 was initiated by Nextremity Solutions on Feb 10, 2022 and is designated Class II. Reason for recall: Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery The recall status is terminated (terminated Oct 10, 2024). Affected quantity: 89 units.

Recall details

Product Type
Devices
Report Date
Apr 20, 2022
Initiation Date
Feb 10, 2022
Termination Date
Oct 10, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
89 units

Product description

Nextra Hammertoe Correction System 3.5 Middle and 3.2 Proximal Saw Cut SKU: NX-3532K-SC Orthopedic surgical instrument

Reason for recall

Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery

Distribution pattern

Distribution US Nationwide, Japan, and the Netherlands

Code information

Lot Number: 168125017B UDI: 00817701020059

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0913-2022
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Nextremity Solutions? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0913-2022. Report a data issue: hello@regdatalab.com.