Z-0739-2023 Class II Ongoing

Recalled by Abbott Molecular, Inc. — Des Plaines, IL

FDA device recall Z-0739-2023 was initiated by Abbott Molecular, Inc. on Nov 22, 2022 and is designated Class II. Reason for recall: Reports of an increase in reactive negative controls and false positive results with certain lots of Alinity M Resp-4-Plex AMP kits. The recall status is ongoing. Affected quantity: 3896 kits.

Recall details

Product Type
Devices
Report Date
Dec 28, 2022
Initiation Date
Nov 22, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3896 kits

Product description

Abbott Alinity m Resp-4-Plex Amp Kit, Part Number 09N79-096

Reason for recall

Reports of an increase in reactive negative controls and false positive results with certain lots of Alinity M Resp-4-Plex AMP kits.

Distribution pattern

Worldwide

Code information

Lot Numbers: 381465, exp 8/24/2023; 524433, exp 1/5/2023; 526710, exp 1/21/2023; 527845, exp 2/8/20231499

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0739-2023
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Abbott Molecular, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0739-2023. Report a data issue: hello@regdatalab.com.