Z-0741-2021 Class II Terminated

Recalled by Westmed, Inc. — Tucson, AZ

FDA device recall Z-0741-2021 was initiated by Westmed, Inc. on Nov 11, 2020 and is designated Class II. Reason for recall: There is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter… The recall status is terminated (terminated Feb 7, 2023). Affected quantity: 3000 units.

Recall details

Product Type
Devices
Report Date
Jan 6, 2021
Initiation Date
Nov 11, 2020
Termination Date
Feb 7, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3000 units

Product description

REF 3102-E, Elongated Oxygen Mask, Adult Oxygen Elongated Mask with 8' Tubing, Rx Only, CE0482, (01)00709078003172 - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.

Reason for recall

There is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter resulting in a possible delay in treatment.

Distribution pattern

Worldwide distribution - US Nationwide distribution including in the states of PA, OK, MO, CO, RI, IL, CA, TX, OH, MD, IL, MN, KS and the countries of Canada, Korea, United Kingdom

Code information

Lot # 091120N53

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0741-2021
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0741-2021. Report a data issue: hello@regdatalab.com.