Z-2435-2020 Class II Terminated

Recalled by Westmed, Inc. — Tucson, AZ

Recall Details

Product Type
Devices
Report Date
July 8, 2020
Initiation Date
January 15, 2019
Termination Date
February 5, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
35 cases

Product Description

Neo-Vent Infant T-Piece Resuscitation Circuit With Infant Mask; Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0420, Qty:10, Length: 60" (1.5cm), Rx Only.

Reason for Recall

A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of CA, GA, IL,OH, MA, MI, MO, NC, OH, OR, TX, VA, and WI. The country of Spain.

Code Information

101618U02, 102318U06, 101518U03.