Z-2435-2020 Class II Terminated

Recalled by Westmed, Inc. — Tucson, AZ

FDA device recall Z-2435-2020 was initiated by Westmed, Inc. on Jan 15, 2019 and is designated Class II. Reason for recall: A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators… The recall status is terminated (terminated Feb 5, 2021). Affected quantity: 35 cases.

Recall details

Product Type
Devices
Report Date
Jul 8, 2020
Initiation Date
Jan 15, 2019
Termination Date
Feb 5, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
35 cases

Product description

Neo-Vent Infant T-Piece Resuscitation Circuit With Infant Mask; Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0420, Qty:10, Length: 60" (1.5cm), Rx Only.

Reason for recall

A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.

Distribution pattern

Worldwide distribution - US Nationwide distribution including in the states of CA, GA, IL,OH, MA, MI, MO, NC, OH, OR, TX, VA, and WI. The country of Spain.

Code information

101618U02, 102318U06, 101518U03.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-2435-2020
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-2435-2020. Report a data issue: hello@regdatalab.com.