Z-0745-2020 Class III Terminated

Recalled by Tosoh Smd Inc — Grove City, OH

Recall Details

Product Type
Devices
Report Date
January 1, 2020
Initiation Date
November 15, 2019
Termination Date
July 1, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1012 units

Product Description

AIA-PACK B12 Pretreatment Set- IVD for diagnostic use for pre-treatment of calibrators/controls. Part Number: 020706

Reason for Recall

The AIA-2000 bar code scanners depend on the quiet zone to help identify the bar code. The lack of an adequate quiet zone can cause the bar code scanner to confuse the bar codes with text or graphics resulting in the scanner ignoring a valid symbol. This results in an error in reading the barcode: Error Code 2220 Unable to read barcode on reagent bottle. In some cases, the quiet zone is too short and an error in barcode reading occurs. The problem is intermittent and not all vials (6 vials are available per set) have been reported to fail. The only risk to patients or users is a delay in results. The company recommends (optional) that customers pre-screen the vials to identify affected products. Customers can contact the company for reimbursement of any affected products. The company has implemented barcode checks as a temporary solution and will replace the barcode printer on the production line for products beginning with lot B*01898.

Distribution Pattern

Nationwide Foreign: Columbia, Dominic Republic, Ecuador

Code Information

Lot Numbers: IZ01893 Exp. 2019-07-31; I801890 Exp. 2019-11-30