Z-1245-2017 Class II Terminated

Recalled by Tosoh Smd Inc — Grove City, OH

FDA device recall Z-1245-2017 was initiated by Tosoh Smd Inc on Dec 22, 2016 and is designated Class II. Reason for recall: Power outages causes reporting software to shutdown. The recall status is terminated (terminated Jun 18, 2019). Affected quantity: 34.

Recall details

Product Type
Devices
Report Date
Mar 1, 2017
Initiation Date
Dec 22, 2016
Termination Date
Jun 18, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
34

Product description

Keyspan High-High Speed USB to Serial Adapter Product Usage: The Reporting Software is an application which serves as middle ware between the Tosoh Automated HPLC G8 analyzer and an LIS or as a stand alone data repository for the analyzer. It is designed to store and manage the received data and make reports as necessary. The application is installed on Dell All- in-One computers that do not have serial (RS232) ports only USB ports. The G8 analyzer does not have USB ports. The Key Span Adaptor connects the serial port on the G8 to the USB port on the Dell computer. Once installed the Keyspan adapter is assigned a COM port. The default for the firmware is dynamic which randomly assigns a port for the transmission of data. The port assignment can change when the power for the analyzer or computer is shut down or the cable is physically disconnected and then reconnected. The default can be changed to geographic which allows the port assignment to be static regardless of a power outage.

Reason for recall

Power outages causes reporting software to shutdown.

Distribution pattern

US Nationwide Distribution in the states of CT, MA, ME, MN, MO, NC, NE, NH, PA, SD, TX, VA, & WV.

Code information

All G8's with Reporting Software interfaced with LIS.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1245-2017
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1245-2017. Report a data issue: hello@regdatalab.com.