Z-0747-2017 Class II Terminated
FDA device recall Z-0747-2017 was initiated by PerkinElmer, Inc on June 24, 2016 and is designated Class II. Reason for recall: Filter Tips molded incorrectly and may not seal to the Varispan tip adaptor properly causing inaccurate aspiration and dispensing The recall status is terminated (terminated July 31, 2018). Affected quantity: 595 trays (571,200 tips).
Recall Details
- Product Type
- Devices
- Report Date
- December 21, 2016
- Initiation Date
- June 24, 2016
- Termination Date
- July 31, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 595 trays (571,200 tips)
Product Description
Perkin Elmer RoboRack 25ul, Clear, Non-Conductive, Filter Tips, Pre-sterilized Item Number: 6000689. Supplied for use with JANUS Automated Workstations: Product Number: AJS4001 AJM4001 AJI4001, AJL4001, AJS8001 , AJM8001, AJI8001, AJL8001, AJM4G01 , AJI4G01 , AJL4G01, AJM8G01, AJI8G01, AJL8G01, AJM4M01, AJI4M01 , AJL4M01, AJM8M01, AJI8M01, AJL8M01, AGS4NGS The JANUS Automated Workstation is an automated, programmable liquid handling instrument intended for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation work flows. The NGS Express is an automated, programmable liquid handling instrument suitable for use in clinical settings to facilitate the pipetting and dilution of solutions in sample preparation workflows
Reason for Recall
Filter Tips molded incorrectly and may not seal to the Varispan tip adaptor properly causing inaccurate aspiration and dispensing
Distribution Pattern
Worldwide Distribution - US including CA, MA, NY, IA, OR , PA, WA and Internationally to Belgium, Canada, Finland, France, Germany, Japan, Netherlands, Singapore, and UK
Code Information
Lot Numbers:15470394, 15471775, 1601067, 16071147