Z-0635-2016 Class II Terminated

Recalled by Perkinelmer — Waltham, MA

FDA device recall Z-0635-2016 was initiated by Perkinelmer on Dec 8, 2015 and is designated Class II. Reason for recall: The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement… The recall status is terminated (terminated Aug 4, 2017). Affected quantity: 47 units.

Recall details

Product Type
Devices
Report Date
Jan 20, 2016
Initiation Date
Dec 8, 2015
Termination Date
Aug 4, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
47 units

Product description

Perkin Elmer WIZARD2 10-detector, 550 samples Product Code: 2470-0100, 3470-0100. The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples

Reason for recall

The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results.

Distribution pattern

Worldwide Distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, MYANMAR, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, and THAILAND.

Code information

Model Number: 2470-0100 Serial Numbers: DG06106436 DG08118240 DG02106050 DG06106502 DG11118609 DG01128795 DG02117461 DG12118702 DG09106861 DG05117833 DG05117834 DG10107006 DG06106452 DG09118367 DG03106154 DG04129081 DG06106559 DG06117903 DG07106599 DG09106774 DG05106387 DG06106524 DG10106949 DG09118284 DG05117787 DG04117755 DG09106770 DG03106196 DG04129159 DG08118192 DG06106469 DG08118189 DG04129075 DG12107252 DG12118677 DG06117923 DG08106741 DG03106186 DG12118759 DG01117444 DG02106112 DG12107236 DG11107031 DG12107249 DG11118612 DG05117781 Model: 3470-0100 (1 unit) Serial Number: DG02128911

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0635-2016
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Perkinelmer? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0635-2016. Report a data issue: hello@regdatalab.com.