Z-0756-2014 Class II Terminated

Recalled by The Anspach Effort, Inc. — Palm Beach Gardens, FL

FDA device recall Z-0756-2014 was initiated by The Anspach Effort, Inc. on Mar 29, 2012 and is designated Class II. Reason for recall: The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be… The recall status is terminated (terminated Mar 11, 2015). Affected quantity: 76.

Recall Details

Product Type
Devices
Report Date
Jan 22, 2014
Initiation Date
Mar 29, 2012
Termination Date
Mar 11, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
76

Product Description

MCA5-1SB - 1MM Fluted Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.

Reason for Recall

The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide

Distribution Pattern

Worldwide Distribution: US (nationwide) and Internationally to: AU, BE, CA, CH, ES, FR, GB, IT, JP, NL, SE, and ZA.

Code Information

All lots are included in recall

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0756-2014
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.