Z-0783-2014 Class II Terminated

Recalled by Zimmer, Inc. — Warsaw, IN

FDA device recall Z-0783-2014 was initiated by Zimmer, Inc. on Dec 19, 2013 and is designated Class II. Reason for recall: After evaluation of 13 complaints(0.4% rate) for hyper-extension of the Zimmer Segmental System, Zimmer is updating the related surgical technique and instructions for use to provide additional… The recall status is terminated (terminated Feb 24, 2015). Affected quantity: 5305 distribution events.

Recall Details

Product Type
Devices
Report Date
Jan 29, 2014
Initiation Date
Dec 19, 2013
Termination Date
Feb 24, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5305 distribution events

Product Description

Zimmer Segmental Systems; Polyethylene Inserts, Size B. Designed to address significant bone loss resulting from oncology, trauma, and/or salvage of previously failed arthroplasty.

Reason for Recall

After evaluation of 13 complaints(0.4% rate) for hyper-extension of the Zimmer Segmental System, Zimmer is updating the related surgical technique and instructions for use to provide additional instructions and patient conditions that may place excessive loading on the polyethylene insert.

Distribution Pattern

Worldwide Distribution. USA nationwide and Germany, with further distribution into EU countries.

Code Information

Part numbers 00-5850-012-95

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0783-2014
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.