Z-0789-2017 Class II Terminated

Recalled by Smiths Medical ASD Inc. — Minneapolis, MN

FDA device recall Z-0789-2017 was initiated by Smiths Medical ASD Inc. on Nov 21, 2016 and is designated Class II. Reason for recall: Smiths Medical became aware that three (3) lots of Acuvance2 Safety Catheter, product number 1720-AI, contain devices with missing or incorrectly positioned side port valves. The recall status is terminated (terminated Aug 1, 2017). Affected quantity: 26,750.

Recall Details

Product Type
Devices
Report Date
Dec 28, 2016
Initiation Date
Nov 21, 2016
Termination Date
Aug 1, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
26,750

Product Description

Acuvance¿2 Safety Catheter, Product No. 1720-AI Product Usage: The Acuvance¿ 2 Safety Catheter is a Peripheral Intravenous Catheter (PIVC). PIVCs are designed for single use venipuncture on all patient populations where access to the veins and arteries are required. The catheter has a Safety Needle Assembly (SAN) designed to reduce risk of accidental needle sticks to the operators. The lateral port is equipped with a non-return silicone valve and a polypropylene snap cap with a winged polypropylene hub for fastening.

Reason for Recall

Smiths Medical became aware that three (3) lots of Acuvance2 Safety Catheter, product number 1720-AI, contain devices with missing or incorrectly positioned side port valves.

Distribution Pattern

Foreign Distribution in the Italy, France, Switzerland, Germany, Sweden

Code Information

3120882, 3120883, 3220733

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0789-2017
openFDA file date
Oct 23, 2026