Z-0793-2020 Class II Ongoing

Recalled by Baxter Healthcare Corporation — Deerfield, IL

FDA device recall Z-0793-2020 was initiated by Baxter Healthcare Corporation on Dec 12, 2019 and is designated Class II. Reason for recall: Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the… The recall status is ongoing. Affected quantity: 271,014,984 units.

Recall Details

Product Type
Devices
Report Date
Jan 15, 2020
Initiation Date
Dec 12, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
271,014,984 units

Product Description

Clearlink System Continu-Flo Administration Sets, Product Codes: 2C8519, 2C8537, and 2C8541 - Product Usage:For the administration of fluids from a container into the patients vascular system through a vascular access device.

Reason for Recall

Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector.

Distribution Pattern

Worldwide distribution - US Nationwide distribution and countries of Bermuda, Bahamas, Trinidad and Tobago, Grand Cayman, St. Thomas, Guyana and Barbados, Mexico, Canada, and New Zealand.

Code Information

Product Codes: 2C8519, UDI 00085412048994; 2C8537, UDI 00085412048970; and 2C8541, UDI 00085412071107 All Lots

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0793-2020
openFDA file date
Oct 26, 2026
Loaded by RegDataLab
Oct 27, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.