Z-0794-2017 Class II Terminated
FDA device recall Z-0794-2017 was initiated by Oscor, Inc. on Oct 5, 2016 and is designated Class II. Reason for recall: Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use. The recall status is terminated (terminated Jul 26, 2019). Affected quantity: 991.
Recall Details
- Product Type
- Devices
- Report Date
- Dec 28, 2016
- Initiation Date
- Oct 5, 2016
- Termination Date
- Jul 26, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 991
Product Description
OSCOR ADELANTE BREEZEWAY 8F C55¿ S79CM D84CM Catalog # AB081044 Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.
Reason for Recall
Complaint that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use.
Distribution Pattern
Worldwide Distribution - US including CA, NJ, MD, TX, FL, NC, MO, and Puerto Rico. Internationally to AUSTRIA, BOLIVIA, CANADA, CHILE, CYPRUS, NETHERLANS, EGYPT, GERMANY, HONG KONG, ISRAEL, ITALY, SWITZERLAND, and VIETNAM
Code Information
Lot #'s: C1-12152 Exp. November 30, 2018, C1-09518 Exp. December 21, 2017, C1-09748 Exp. February 16, 2018, C1-09899 Exp. March 12, 2018, C1-10040 Exp. April 5, 2018, C1-10040 Exp. April 5, 2018, C1-10149 Exp. April 23, 2018, C1-10330 Exp. May 11, 2018, C1-10329 Exp. May 11, 2018, C1-10382 Exp. May 17, 2018, C1-11571 Exp. September 25, 2018, C1-11571 Exp. September 25, 2018, and OR-04757 Exp. December 22, 2018