Z-0821-2024 Class II Ongoing
FDA device recall Z-0821-2024 was initiated by American Contract Systems, Inc. on Nov 29, 2023 and is designated Class II. Reason for recall: During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating… The recall status is ongoing. Affected quantity: 95 units.
Recall Details
- Product Type
- Devices
- Report Date
- Feb 7, 2024
- Initiation Date
- Nov 29, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 95 units
Product Description
Custom procedural convenience kits and trays, gastroenterological & urological, labeled as: AHN BIOPSY STERILE, kit number AHBP12A
Reason for Recall
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
Distribution Pattern
US Nationwide distribution.
Code Information
AHBP12A, UDI/DI 00191072188229, Lot Numbers: 963231, exp. 1/21/2024; 940231, exp. 1/21/2024; 996231, exp. 2/28/2023
Source
- Dataset