Z-0828-2022 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- April 6, 2022
- Initiation Date
- November 16, 2021
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 14 units
Product Description
cobas LiatSystem, respiratory virus panel nucleic acid assay system, Material Number 07341920190
Reason for Recall
Roche has identified a small number of cobas Liat analyzer units that have consistently high noise levels (background signal) originating from the amber detector (used for Influenza B detection). This noise increases the probability of false positive Influenza B results.
Distribution Pattern
Worldwide distribution - US Nationwide distribution in the states of CO, IL, IN, KY, MI, NC, and TX. The countries of Cyprus, Spain, and Hong Kong.
Code Information
UDI: 07613336100097; Serial Numbers: 16333, 16454, 16688, 16695, 16871, 16914, 17048, 17266, 19924, 18094, 18323, 18357, 17116, 17086