Z-0828-2022 Class II Ongoing

Recalled by Roche Molecular Systems, Inc. — Branchburg, NJ

Recall Details

Product Type
Devices
Report Date
April 6, 2022
Initiation Date
November 16, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14 units

Product Description

cobas LiatSystem, respiratory virus panel nucleic acid assay system, Material Number 07341920190

Reason for Recall

Roche has identified a small number of cobas Liat analyzer units that have consistently high noise levels (background signal) originating from the amber detector (used for Influenza B detection). This noise increases the probability of false positive Influenza B results.

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of CO, IL, IN, KY, MI, NC, and TX. The countries of Cyprus, Spain, and Hong Kong.

Code Information

UDI: 07613336100097; Serial Numbers: 16333, 16454, 16688, 16695, 16871, 16914, 17048, 17266, 19924, 18094, 18323, 18357, 17116, 17086