Z-1106-2023 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- February 15, 2023
- Initiation Date
- December 21, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 727 kits OUS only
Product Description
cobas Influenza A/B & RSV UC (Utility Channel) Qualitative nucleic acid test for use on the cobas 6800/8800 Systems; Material number 09233962190
Reason for Recall
The firm received customer complaints regarding false negative Influenza A (Flu A) results and late Flu A Target Ct values. A false negative influenza A result may lead to additional testing, psychological distress, and delays to targeted therapy for influenza as well as true diagnosis.
Distribution Pattern
Domestic distribution nationwide. Foreign distribution worldwide.
Code Information
UDI-DI 00875197006773 All in-date lots