Z-0828-2024 Class II Ongoing

Recalled by American Contract Systems, Inc. — Temple Terrace, FL

FDA device recall Z-0828-2024 was initiated by American Contract Systems, Inc. on Nov 29, 2023 and is designated Class II. Reason for recall: During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating… The recall status is ongoing. Affected quantity: 1020 units.

Recall Details

Product Type
Devices
Report Date
Feb 7, 2024
Initiation Date
Nov 29, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1020 units

Product Description

Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LOWER PELVIC PACK, kit number AGLP52L; b) OB/GYN DAVINCI PACK A RG BASIN, kit number HGOB98G; c) DAVINCI GYN PACK, kit number HHDV17; d) KIT, GYNE ROBOTIC, kit number MMGR44D; e) KIT, MAJOR LITHOTOMY, kit number MMJL16E; f) GYN PACK, kit number SMGY46; g) Lithotomy Pack, kit number THLH71C; h) ASC HYSTEROSCOPY PACK (PS 020634), kit number UIHY16AA; i) GYN LAPAROSCOPY HPES, kit number WELP67C; j) GYN LAPAROSCOPY HPES, kit number WELP67C-01; k) GYN LAPAROSCOPY HPES, kit number WELP67C-02

Reason for Recall

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Distribution Pattern

US Nationwide distribution.

Code Information

a) AGLP52L, UDI/DI 00191072144119, Lot Numbers: 933231, exp. 3/8/2024; 942231, exp. 2/27/2024; 947231, exp. 2/22/2024; 966231, exp. 2/3/2024; 980231, exp. 1/20/2024; 681221, exp. 11/15/2023; 707221, exp. 10/20/2023; b) HGOB98G, UDI/DI none, Lot Numbers: 667221, exp. 11/29/2023; 977231, exp. 1/23/2024; 952231, exp. 1/31/2024; 896231, exp. 1/31/2024; c) HHDV17, UDI/DI 00191072170125, Lot Numbers: 766221, exp. 8/22/2023; 731221, exp. 9/26/2023; 682221, exp. 11/14/2023; 638221, exp. 12/28/2023; 946231, exp. 2/23/2024; d) MMGR44D, UDI/DI none, Lot Numbers: 709221, exp. 10/18/2023; 650221, exp. 12/16/2023; 891231, exp. 4/19/2024; e) MMJL16E, UDI/DI none, Lot Numbers: 989231, exp. 1/11/2024; 920231, exp. 3/21/2024; f) SMGY46, UDI/DI 00191072188458, Lot Numbers: 961231, exp. 2/8/2024; g) THLH71C, UDI/DI 00191072155108, Lot Numbers: 756221, exp. 9/1/2023; 708221, exp. 10/19/2023; h) UIHY16AA, UDI/DI none, Lot Numbers: 931231, exp. 3/10/2024; 896231, exp. 4/14/2024; i) WELP67C, UDI/DI 00191072173676, Lot Numbers: 751221, exp. 9/6/2023; 681221, exp. 9/13/2023; j) WELP67C-01, UDI/DI 00191072173676, Lot Numbers: 720221, exp. 9/21/2023; k) WELP67C-02, UDI/DI 00191072173676, Lot Numbers: 643221, exp. 9/13/2023

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0828-2024
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.