Z-0834-2020 Class II Terminated

Recalled by Polymer Technology Systems, Inc. — Indianapolis, IN

FDA device recall Z-0834-2020 was initiated by Polymer Technology Systems, Inc. on Dec 2, 2019 and is designated Class II. Reason for recall: One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to… The recall status is terminated (terminated Jul 6, 2020). Affected quantity: 44,448.

Recall details

Product Type
Devices
Report Date
Jan 29, 2020
Initiation Date
Dec 2, 2019
Termination Date
Jul 6, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
44,448

Product description

Henry Schein CardioChek Starter Kit, Reference Number 4060 HS, 570-1036

Reason for recall

One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.

Distribution pattern

The products were distributed US Nationwide. The products were distributed to the following foreign countries: Australia, Austria, Belgium, Bermuda, Canada, China, Chile, Costa Rica, Cayman Islands, Czech Republic, England, Finland, France, Germany, Greece, Hong Kong, India, Iraq, Italy, Jordan, Lebanon, Lithuania, Mexico, Morocco, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Trinidad, UAE.

Code information

All Serial Numbers shipped during the 11/1/2016 to 10/25/2019 timeframe. UDI: 00304040040539

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0834-2020
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Polymer Technology Systems, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0834-2020. Report a data issue: hello@regdatalab.com.