Z-1372-2019 Class II Terminated

Recalled by Polymer Technology Systems, Inc. — Indianapolis, IN

Recall Details

Product Type
Devices
Report Date
May 29, 2019
Initiation Date
January 16, 2019
Termination Date
June 24, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
67643 total

Product Description

Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412

Reason for Recall

Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being di P832 Chemistry (Diagnostic) CHH 2 Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412

Distribution Pattern

Domestic distribution nationwide.

Code Information

Q802 Q811 Q813 Q816 Q820 Q823 Q831 Q903