Z-1372-2019 Class II Terminated
FDA device recall Z-1372-2019 was initiated by Polymer Technology Systems, Inc. on January 16, 2019 and is designated Class II. Reason for recall: Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being di P832 Chemistry (Diagnostic) CHH 2 Lipid Pane… The recall status is terminated (terminated June 24, 2020). Affected quantity: 67643 total.
Recall Details
- Product Type
- Devices
- Report Date
- May 29, 2019
- Initiation Date
- January 16, 2019
- Termination Date
- June 24, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 67643 total
Product Description
Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412
Reason for Recall
Some lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being di P832 Chemistry (Diagnostic) CHH 2 Lipid Panel test strips (as a component of smart bundles), REF 2729, also private labeled under Henry Schein as REF 570-0412
Distribution Pattern
Domestic distribution nationwide.
Code Information
Q802 Q811 Q813 Q816 Q820 Q823 Q831 Q903