Z-0839-2021 Class II Ongoing

Recalled by Fresenius Medical Care Holdings, Inc. — Waltham, MA

Recall Details

Product Type
Devices
Report Date
January 20, 2021
Initiation Date
November 23, 2020
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
224 drums

Product Description

Naturalyte Liquid Acid concentrate drums, Product Code 13-2251-0 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

Reason for Recall

The deviation in conductivity for sodium level exceeds 0.6mS/cm from the theoretical conductivity (TCD) or approximately 6mEq/L of sodium from prescriptions that was previously communicated. Calcium concentration is below manufacturing specification.

Distribution Pattern

US Nationwide distribution.

Code Information

Lot Numbers: 20LXAC025 20LXAC081 20LXAC082 20LXAC083 Additional Affected lots as of 3/17/21: 20LXAC029 20NXAC021 20LXAC026 20LXAC065 20LXAC028 20NXAC029 20LXAC071 20LXAC103 20LXAC027 20LXAC070 20LXAC067 20NXAC019 20LXAC082 20LXAC025 20LXAC081 20LXAC083