Z-0849-2015 Class II Terminated

Recalled by Illumina Inc — San Diego, CA

FDA device recall Z-0849-2015 was initiated by Illumina Inc on Nov 13, 2014 and is designated Class II. Reason for recall: It was discovered a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The software cannot report insertions or… The recall status is terminated (terminated Mar 18, 2015). Affected quantity: 13 units.

Recall details

Product Type
Devices
Report Date
Dec 31, 2014
Initiation Date
Nov 13, 2014
Termination Date
Mar 18, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13 units

Product description

Illumina MiSeqDx Universal Kit 1.0, PN 15039608 The Illumina MiSeqDx Universal Kit 1.0 is a set of reagents and consumables used in the processing of human genomic DNA samples derived from peripheral whole blood, and in the subsequent targeted sequencing of the resulting sample libraries. User-supplied analyte specific reagents are required for the preparation of libraries targeting specific genomic regions of interest. The MiSeqDx Universal Kit 1.0 is intended for use with the MiSeqDx instrument.

Reason for recall

It was discovered a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The software cannot report insertions or deletions at specific genomic locations with respect to the primer.

Distribution pattern

US Distribution to the states of AZ, GA, CA, OH, and VA., and one Internationally to France.

Code information

Serial No. M70130, M70189, M70108, M70165, M70166, M70175, M70172, M70103.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0849-2015
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Illumina Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works