Z-0849-2015 Class II Terminated
FDA device recall Z-0849-2015 was initiated by Illumina Inc on Nov 13, 2014 and is designated Class II. Reason for recall: It was discovered a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The software cannot report insertions or… The recall status is terminated (terminated Mar 18, 2015). Affected quantity: 13 units.
Recall details
- Product Type
- Devices
- Report Date
- Dec 31, 2014
- Initiation Date
- Nov 13, 2014
- Termination Date
- Mar 18, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 13 units
Product description
Reason for recall
Distribution pattern
US Distribution to the states of AZ, GA, CA, OH, and VA., and one Internationally to France.